Knauer Instrument Control Method Options 175<br />

Fig. 259 Solvent Control dialog box – HPG system<br />

The amount of solvents used will be calculated correctly only under<br />

software control. Should the pumps have been controlled manually,<br />

or the bottles have been refilled, the actual solvent amount must be<br />

corrected.<br />

Qualification Procedures<br />

ChromGate ® provides a performance qualification tool for checking your<br />

HPLC systems. This is not a qualification procedure for each individual<br />

device (pump, detector etc.) but for the whole configured chromatographic<br />

instrument. The procedure is at last a normal method used to<br />

determine the measurement reproducibility.<br />

The PQ procedure is started from the main instrument screen by<br />

selecting the menu sequence<br />

Fig. 260 PQ start screen<br />

CONTROL – QUALIFICATION PROCEDURES<br />

You have the choice to create a new PQ session method or to select<br />

earlier created ones. These may be completed or not. If you select a not<br />

completed one you can continue with it. Selecting a completed one the<br />

final Report will be displayed.