144 Creating an Instrument Control Method<br />

All direct control settings will also be stored in the Instrument Activity Log.<br />

To see the log, click on “File – Instrument Activity Log – Display Log”.<br />

Fig. 215 Extend run time window<br />

Control Pressure Limits<br />

This section allows you to enter the control pressure limits in the units<br />

chosen in the pump configuration dialog. The pump will automatically<br />

shut down its operation when the measured pressure is outside of the<br />

limits.<br />

Min Pressure<br />

Enter the minimum pressure allowable for your pump.<br />

Max Pressure<br />

Enter the maximum pressure allowable for your pump.<br />

Apply<br />

Click the button to actualize the selected Min/Max pressure limits.<br />

Events<br />

Click the button to set up the event outputs. Refer to the chapter<br />

Instrument Setup – Pumps, section Configuring Events for details. Click<br />

the Events Apply button to accept the settings. This status area is only<br />

shown for pumps offering this feature.<br />

Purge<br />

Fig. 216 Purge settings window<br />

Click the Purge button to open the purge set up window. In Purge flow<br />

(ml/min) enter the desired flow rate. Click on the Start button to start the<br />

pump. In the Gradient components section you can select each solvent<br />

channel directly or a “Mix” (ABCD). You can set a time limit for the purge.<br />

Enable the Stop after option and enter the desired purging time.