Creating an Instrument Control Method 95<br />

Instrument Setup – Kontron Detectors<br />

(3xx, 4xx, and 5xx)<br />

Fig. 138 Kontron Detector setup window<br />

Except the not available option Acquisition Channel On/Off the general<br />

detector settings as described on page 91 can be made. Additionally the<br />

range can be set to define the range of the analog detector outputs. The<br />

choice is 0.100, 0.200, 0.500, 1.000, and 2.000 Absorption Units Full<br />

Scale (AUFS).<br />

The Kontron x30 and x35 detectors can be used in a dual channel mode.<br />

As far as the option Used Channel #2 is checked, a second wavelength<br />

column will appear in the spreadsheet below. Also the otherwise not<br />

accessible cycle time becomes active for changing. This defines the<br />

alternating measuring time for the two channels. The data acquisition is<br />

not performed truly simultaneously. The cycle time you can set within the<br />

range of 0.40 through 1.00 seconds with 0.05 s increments.<br />

For the x32 detectors the option Used Channel #2 is not available and<br />

the cycle time is not displayed at all.<br />

The wavelength table is a spreadsheet to enter your time (detector)<br />

program. In the first row (at time 0.00 min), enter the initial wavelength(s)<br />

in nm. The editing principle is the same as for the time (gradient)<br />

program, described in the chapter Instrument Setup – Pumps, Pump<br />

Program.<br />

Instrument Setup – K-2600 Detector<br />

The detector K-2600 supports the general detector settings described on<br />

page 91.<br />

The appearance of some of the details of the K-2600 detector tab<br />

depends on the availability of a PDA option.<br />