40 Setup and Control of Knauer HPLC Systems<br />

Name<br />

The name of the selected pump appears automatically. You can enter<br />

any other descriptive name for the pump. This name will appear on the<br />

pump tab when the instrument setup or status windows are displayed.<br />

Use a name which is unique within the instrument (configuration).<br />

E.g. for a high-pressure gradient system, all pumps must have<br />

different names.<br />

Pump Type (accessible for Pump K-120/501 only)<br />

Select the type of pump you want to configure from the drop-down list.<br />

The choices are K-120 and K-501.<br />

Fig. 48 selection option for K-120, K-501 pumps<br />

Name / Type (accessible for Pumps 10P/20P/P2.1S/P4.1S only)<br />

Select the type of pump you want to configure from the drop-down list.<br />

The choices are 10P, 20P, P2.1S and P4.1S. The 10P/P2.1S is the<br />

device without the pressure sensor. As already described in the Name<br />

section above, you must use different names for all of the devices in an<br />

instrument (configuration).<br />

Fig. 49 selection option for pumps 10P, 20P, P2.1S and P4.1S<br />

Gradient Mode<br />

Fig. 50 selection options for gradient modes<br />

Select the desired gradient mode from the drop-down list. Please note,<br />

that for pumps K-1000/1, S 1000, S 1050, preparative pump 1800/K-1800<br />

and P2.1L the gradient mode must correspond with the gradient mode<br />

set in the pump. The LPG option is not available for S<br />

100/10P/20P/P2.1S/P4.1S, K-120 and K-501 pumps. The Isobar mode is<br />

only available for the pump P2.1L.<br />

Interface / Serial Port<br />

Select from the drop-down list the serial port or network number for the<br />

communication port on your PC where the instrument is connected.<br />

To ensure an error-free operation, each serial-controlled device<br />

should be connected with an own COM port, also for devices that<br />

allow to connect more than one device onto the same COM port.<br />

For pumps K-1000, K-1001, S 1000 and K-1800 (preparative pump 1800)<br />

the interface “Network” means the serial port number.<br />

The pumps K-1000, K-1001, S 1000 and K-1800 (preparative pump<br />

1800) only are controlled with ChromGate, if the pump’s internal