SEC Option 277<br />

Fit Type<br />

Select a calibration curve fit type from the drop-down list. The choices<br />

include: Point-to-point, Linear, Quadratic, Cubic, 4th Degree Polynomial,<br />

and Spline.<br />

Goodness of Fit<br />

The calibration curve for the data in the SEC Setup will be displayed in<br />

the calibration curve window. The r 2 value, which represents goodness<br />

of fit of the calibration points to the curve selected, will be displayed<br />

above the curve. A value of 1.00 indicates perfect fit. Select the Time or<br />

Volume button at the bottom of the calibration curve to display the x-axis<br />

of the curve as either retention time in minutes or retention volume. If<br />

you select volume, enter the flow rate in ml/min.<br />

Remove calibration curve outliers<br />

To temporarily remove an outlier from the calibration curve, click on the<br />

point with your mouse. It will change color and the curve will be recalculated<br />

with the point removed.<br />

Time/Volume<br />

Select the mode of viewing for your calibration curve: Time or Volume. If<br />

you select Volume, you must enter the flow rate so that the curve is<br />

displayed correctly.<br />

Print<br />

Click this button to print your current SEC method information.<br />

Export settings<br />

Click this button to select export options and path.<br />

When you have completed the SEC Setup, you are ready to acquire and<br />

process your SEC runs. You can acquire data either manually (single<br />

injection), or by batch acquisition (auto injector). See the ChromGate ®<br />

User’s Guide for information on creating a batch sequence.<br />

Zooming on the Calibration curve<br />

You can zoom on the SEC Calibration curve by clicking and dragging<br />

with the mouse. Double-click the mouse in the graph area to unzoom the<br />

curve view.<br />

Once you have zoomed in on a chromatogram, you can scroll the<br />

chromatogram to the right or left without losing the zoom. This is done by<br />

pressing the keys down and moving the mouse until<br />

the cursor changes to a "hand" and dragging it to the left or right.<br />

You can also scroll the X- or Y- axis to view features which may be out of<br />

the range. To do this, press the keys down while the<br />

mouse cursor is outside the graph area, yet near the axis of interest. The<br />

cursor will change to an up/down arrow near the Y-axis, or a left/right<br />

arrow near the X-axis. Moving the mouse in this mode will scroll the<br />

graph up/down or left/right on the axis.<br />

To restore the original view, do a right-hand mouse click in the<br />

chromatogram window, followed by the Full Unzoom command.<br />

Removing Calibration Outliers<br />

You can temporarily remove an outlier from your calibration curve by<br />

clicking on the outlier point on the calibration curve. When you select an<br />

outlier point this way, the point will turn red, as will its Molecular Mass