70 Setup and Control of Knauer HPLC Systems<br />

Configuration – Autosampler 3800<br />

Smartline Autosampler 3800 and K-3800 (Basic Marathon)<br />

Fig. 98 Configuration window for the autosampler 3800<br />

Name<br />

Enter a descriptive name for the sampler. This name will appear on the<br />

autosampler tab when the instrument setup or status windows are<br />

displayed. Use a name which is unique within the instrument.<br />

Device ID<br />

Select the device identifier for the autosampler from the drop-down list.<br />

This must match the settings of the autosampler being used. To check<br />

the device identifier for the Smartline Autosampler 3800 or the K-3800<br />

(Basic Marathon) press the PROG/END button on the autosampler<br />

keyboard once in serial mode.<br />

Serial Port<br />

Select the number for the communication port on your PC where the<br />

autosampler is connected.<br />

Serial Number<br />

Type in the serial number of the detector you are using.<br />

Loop Volume<br />

Enter the volume of the installed loop in µl.<br />

Sample Tray<br />

Select the type of the installed sample tray. The choices are 96 vials of<br />

1.5 ml and 48 vials of 5.0 ml.<br />

Tray cooling<br />

Select this box if you have a built-in Peltier cooling option installed and<br />

want to cool down your samples. The temperature setting will be done<br />

within the instrument setup.<br />

Partial loop fill<br />

Check this box to select the partial loop fill injection mode (for the Basic<br />

Marathon Plus only).<br />

Add. Info<br />

Click the button Add. Info... to enter a more detailed description of the<br />

autosampler. Enter additional information or comments to be printed out<br />

when you print the instrument configuration.<br />

When complete, click OK to exit the dialog and return to the instrument<br />

configuration icon list.<br />

The autosampler 3800 must be set in to serial mode to be controlled<br />

by software. Press the buttons and on the autosamplers<br />

keypad, the display will show then “Serial Mode Press F0 to exit”.