Knauer Instrument Control Method Options 171<br />

<br />

If you click the button, three additional options are<br />

available.<br />

Fig. 250 Knauer Instrument Control Method Option - <strong>Advanced</strong> Runtime Settings<br />

Check run time when saving method<br />

ChromGate allows different run times for all devices. If you wants to have<br />

the same run time for all devices and let the software check if, e.g. after a<br />

change, all devices have the same run time, enable this option. If there<br />

are different run times, the software displays an error message with the<br />

run time of all devices. The run times must be equalized before you can<br />

save the message.<br />

Fig. 251 Error message Run time check failed<br />

If enabled, this option is valid for all instruments of the Enterprise; it<br />

is not limited on the currently used instrument or computer. That<br />

means, that the run time will checked for all instruments and<br />

therefore in all methods that will be stored on the mentioned<br />

Enterprise machine.<br />

System wake-up and all lamps on<br />

If the option “System wake-up and all lamps on” is enabled and a date<br />

and time is selected, all device, switched to Standby and the lamps of all<br />

connected detectors will be switched on at the selected time. The method<br />

must be opened and the detectors switched on or in Standby at this time.<br />

The wake-up date time you can set by direct editing the time in the dateand-time<br />

line. If you click on the arrow key , you can select a date from<br />

a monthly overview (refer to the picture below).<br />

Please note, that Standby only works for all Azura and the newer<br />

Smartline devices pump S 1050 and detectors S 2550 und S 2520. Older<br />

Smartline devices as pump S 1000, detector S 2600 cannot wake-up<br />

from Standby by a software command; also the lamp of the detector<br />

cannot be switched on, if the detector is in Standby.