ChromGate® System Suitability Setup 187<br />

2. Select a method for calculation of the performance values by clicking<br />

on the Calculation Method down-arrow.<br />

3. Choices include USP, EMG (Exponential Modified Gaussian), DAB,<br />

JP, and AOH (Area/Height) method. The calculations used are given<br />

in the equations section. For BP, EP, and ASTM methods, choose<br />

DAB.<br />

4. Click on the Calculate Performance Parameters box to enable the<br />

suitability calculations for the method<br />

Running a Suitability Test<br />

Fig. 280<br />

Once you have completed the System Suitability Setup and<br />

Performance dialogs, create a Sequence or modify an existing<br />

Sequence to include one or more system suitability runs according to the<br />

needs and SOP’s of your laboratory. If you want the System Suitability<br />

Report to print automatically at the end of the sequence, you must also<br />

select the Print Sequence Reports option from the Sequence<br />

Properties dialog box.<br />

For example, following USP standards, five replicate standards are run at<br />

the beginning of the sequence. These are designated as system<br />

suitability standards in the sequence. At the end of the sequence,<br />

suitability calculations are made, and EZChrom Elite Client/Server<br />

generates a system suitability report.<br />

Run Type is selected by clicking on the Run Type field, then selecting<br />

the appropriate sample type from the choices provided.