Creating an Instrument Control Method 165<br />

[>]<br />

Pressing the [>]-button opens a window, that allows you for perform<br />

several service operations.<br />

Tray position<br />

You can move the tray to several positions. The choices are Tray<br />

Home, Needle exchange position and Tray Front.<br />

Syringe position<br />

The syringe can be moved to Home, End or Exchange position.<br />

Valve position<br />

The valve can be switch to Inject and Load position.<br />

Instrument Status – Triathlon/Endurance Autosampler<br />

Fig. 242 Autosampler Triathlon status tab<br />

Monitor<br />

AS Status<br />

The status Ready will be displayed when the autosampler is waiting to<br />

perform an injection. If an injection program is running, the actual<br />

program step will be indicated. After a completed injection, the status is<br />

described as Analysis timer running during the chromatographic run.<br />

Current Vial<br />

The actual position of the injection needle is shown.<br />

Tray Temperature (°C)<br />

If the tray cooling option has been installed and configured the actual tray<br />

temperature will be shown.<br />

Direct Control<br />

The direct control features are not accessible during an injection.<br />

Tray Temperature (°C)<br />

If the tray cooling option has been installed and configured the tray<br />

temperature can be set. The entered value will be transferred to the<br />

sampler by clicking the Apply button. The On / Off button can be used<br />

for enabling / disabling the cooling control. The labeling of the button<br />

changes with the actual status.<br />

Start Initial Wash<br />

Press this button to force the autosampler to perform the needle wash.