12.02.2013 Views

ChromeGate 3.3.2 Software Manual - KNAUER Advanced Scientific ...

ChromeGate 3.3.2 Software Manual - KNAUER Advanced Scientific ...

ChromeGate 3.3.2 Software Manual - KNAUER Advanced Scientific ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ChromGate® System Suitability Setup 185<br />

ChromGate ® System Suitability Setup<br />

Fig. 278<br />

Before ChromGate ® can make System Suitability calculations, you must<br />

enter the required acceptance ranges for the peaks of interest. This is<br />

done in using the Method – System Suitability command.<br />

1. Click on the Method – System Suitability command to view the<br />

System Suitability Setup dialog.<br />

2. Select the first peak to be used for calculations by highlighting it with<br />

the mouse in the Compounds list.<br />

3. In the adjacent spreadsheet, click on the first field in the Parameter<br />

column. A drop-down list of available parameters is displayed. Select<br />

a parameter from the list. If you do not wish to perform a suitability<br />

test on any given peak, simply leave the Parameter fields blank.<br />

Similarly, if you do not wish to have one of the test criteria used (for<br />

example, %RSD), leave it blank. Some parameters have a choice of<br />

calculation methods. These parameters have the calculation method<br />

displayed after the parameters in parenthesis (e.g. Plates/Meter (JP)<br />

indicates Plates/Meter calculated using the Japanese Pharmacopeia<br />

calculation). For details on these calculations, see the equations<br />

section.<br />

4. For each parameter selected, enter a minimum value (Min),<br />

maximum value (Max), and maximum allowed percent relative<br />

standard deviation (%RSD).<br />

5. If you want a noise test to be performed, click on the lower<br />

spreadsheet in the column labeled Test. Select one of the noise<br />

calculations Noise (rms)/ASTM Noise (rms), or 6-sigma Noise<br />

(rms) from the drop-down list. If you select this box, you must enter a<br />

Start and Stop time for the test, and the Threshold value to<br />

determine acceptable limit. The RMS noise value for the portion of<br />

chromatogram between the Start and Stop times will be calculated<br />

and compared to the Threshold value to determine whether the test<br />

Passed or Failed. Note that the times you enter for the noise test<br />

should be representative of a baseline area of your chromatogram<br />

where not peaks elute. Keep in mind, that the noise may depend on<br />

the detector type, wave length, flow rate, solvent and other<br />

parameters used in the test.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!