ChromGate® Preparative Option 243<br />

Use as reference for correction next events<br />

Normally any shift in retention times is increasing with the<br />

magnitude of the retention. Using the reference option ChromGate<br />

adapts the shift for the calculation of the expected (allowed)<br />

interval. It is possible to set more than one event as reference.<br />

According to the elapsed run time always the last one will be used<br />

for calculation.<br />

Fig. 344 FC parameters Use as reference (single event)<br />

After closing the window the parameter indication will be<br />

supplemented by (Ref).<br />

Action selection: None<br />

Selecting this action option neither the recycling valve nor the<br />

fractionation valve will be switched. The previous action is<br />

continued, only a marker is set. This is useful especially in<br />

combination with the option relative to the previous event. Thus for<br />

instance a collection can be started at a defined time after the set<br />

conditions are met.<br />

Fraction Position<br />

This field is only informative according to the settings in the parameter<br />

window. Either the defined vial position is displayed or next for the first<br />

free vial position. Further possibilities are Waste, Peak recycling,<br />

Solvent Recycling, and in case of action None 0.00 is displayed. In<br />

case of Collect Slices always next is shown.<br />

Fraction Check<br />

If this option is checked any run will be interrupted if the defined<br />

fractionation conditions will not met. It is not possible, directly to<br />

continue such an interrupted run. In case of peak recognition mode<br />

this option cannot be used.<br />

Rack configuration path<br />

In any method the fraction collector program must connected to a rack<br />

configuration. The rack configuration file contains information about<br />

collected fractions in one or several runs. Opening the fraction collector<br />

tab of the setup window first time, no rack configuration path will be<br />

displayed.