66 Setup and Control of Knauer HPLC Systems<br />

Fig. 94 Tray configuration 84+3 vials for the autosampler 3950<br />

The “84+3 vial” tray for the autosampler 3950 is supported in<br />

firmware version 1.14 or higher. A firmware upgrade from former<br />

firmware versions is not available.<br />

Left (Right) tray type<br />

Select the type of the installed tray for both sides. The choices are<br />

96 low, 96 high, 384 low, 48 vials, and 12 vials. The selection can<br />

be different for each tray side. In case of activated prep mode it is<br />

fixed on 12 vials. If the tray type 84+3 vials is enabled, this choice is<br />

for both side; this tray uses the left and right tray side. Only for the tray<br />

type 84+3 vials it is possible to select a source vial for the transport<br />

liquid in the method setup later. The 84+3 vials tray is only available<br />

for autosamplers with firmware version 1.14 or higher. Autosamplers<br />

with a previous firmware version cannot be updated.<br />

Plate process order<br />

The vials of the two trays will be processed either in columns or rows<br />

order.<br />

First Destination Vial<br />

Enable this option if you want to use destination vials. Enter the first<br />

destination vial position. For each injection a destination vial will be<br />

used for mixing a diluting.<br />

Reagent A(B) Vial<br />

Enable these options if you want to use reagent vials. Enter the<br />

corresponding vial position. The selected vial will be used for Reagent<br />

A and/or Reagent B for all injections.<br />

Use Mix & Dilute<br />

Enable this option to select for each injection one destination, reagent<br />

A and reagent B vial (position). This allows a higher number of<br />

injections if a higher volume of the reagent(s) is required.<br />

The Mix & Dilute option is only available if you have selected the<br />

same tray type for left and right. It is not available for the tray type<br />

84+3 vials.