98 Creating an Instrument Control Method<br />

Instrument Setup – S 2550 Detector<br />

The detector S 2550 supports the general detector settings described on<br />

page 91.<br />

Fig. 143 Detector S 2550 setup window<br />

Used Channel<br />

Check the box(es) to turn the acquisition for the selected channel(s) on.<br />

Only channels that have been selected in the instrument configuration of<br />

the detector are available. If the box is not selected, no data will be<br />

acquired and no time program will be executed for this channel. As many<br />

channels are activated (at least one) as many wavelength columns<br />

appear in the spreadsheet below.<br />

The wavelength table is a spreadsheet to enter your time (detector)<br />

program. In the first row (at time 0.00 min), enter the initial wavelength(s)<br />

in nm. The editing principle including programming the digital outs is the<br />

same as for the time (gradient) program, described in the chapter<br />

Instrument Setup – Pumps, Pump Program.<br />

Autozero at WL change<br />

Check this option if you wish the signal to be zeroed at the moment the<br />

wavelength is changed (to define the current signal as the baseline).<br />

[Action] at the end of run<br />

The detector S 2550 allows not only to switch off the lamp at the end of<br />

run, but also to switch the device into Standby.<br />

You should switch on the lamp either manually or from the direct<br />

control window before you start the next working session. After<br />

power up, allow 20 minutes for the lamp to stabilize before running<br />

samples that may require sensitive measurements.<br />

Scan, nm [Range xxx – yyy nm]<br />

This section allows you to specify the scan spectral interval, in nm.<br />

Start<br />

Enter the start wavelength for the scan in nm.<br />

End<br />

Enter the end wavelength for the scan in nm.<br />

Acquire 3D data<br />

You have access to this selection field only with the PDA option. The<br />

displayed range depending on the detector type and the installed lamps<br />

will only be updated, if you open that window online. The <strong>Advanced</strong>…<br />

button becomes accessible if the Acquire 3D data check mark is set.