100 Creating an Instrument Control Method<br />

Fig. 146 Detector S 2550 time based scan<br />

Instrument Setup – Diode Array Detectors<br />

(S 2600, DAD 2850, DAD 2800, K-2700 and Kontron DAD 540/545)<br />

The S 2600, S 2850, S 2800, K-2700, and K-2800 detectors are fullyfunctional<br />

DAD (diode array detector) or PDA (photo diode array)<br />

detectors with the capability to simultaneously collect data over the entire<br />

UV and visible spectra.<br />

These detectors support the general detector settings described on page<br />

91.<br />

Fig. 147 Detector S 2600 setup window with PDA option<br />

Options<br />

An. output (no access with Kontron DAD 540)<br />

You can activate the option to select the scaling of the analogue output<br />

signal. This option is not compatible with a set sampling rate of 10 Hz.<br />

The following error message will be displayed when the window will be<br />

closed:<br />

Fig. 148 Error message<br />

Scale<br />

The selected value represents the full scale. Values are selectable from<br />

0.001 AU up to 10 AU. For the Kontron DAD 540 it is fixed on 10 AU<br />

Autozero at WL change<br />

Check this option if you wish the signal to be zeroed at the moment the<br />

wavelength is changed (to define the current signal as the baseline). At<br />

the run start always an Autozero is performed.