154 Creating an Instrument Control Method<br />

Scanning speed<br />

Select a value between 24 and 3000 nm/min from the pull-down menu.<br />

Sensitivity, Gain, Output Mode<br />

Select the sensitivity level (HIGH, MED, or LOW), the gain (1, 4, 16), and<br />

the output mode (Emission NOT Corrected, Excitation Energy, Light<br />

Source Corrected) from the corresponding pull-down menus.<br />

Analog output<br />

An analog output can be activated. The Recorder range is combined<br />

with this option. You can define this range by selecting one from the pulldown<br />

menu.<br />

Scan<br />

A click on this button will start the diagnostic measurement. It will take<br />

some time depending on your scan range and speed settings. During the<br />

scan, the info “please wait” will blink in the head bar. The spectrum will be<br />

displayed after finishing the scan.<br />

The obtained diagnostic spectrum cannot be saved directly or<br />

exported! For later comparisons you must take a screenshot and<br />

save this in an appropriate way.<br />

Instrument Status – Conductivity Detector Alltech 650<br />

Fig. 228 Detector Status tab, Conductivity Detector Alltech 650<br />

Monitor<br />

Run Status<br />

The status HOLD will be displayed when no program is running;<br />

otherwise the Run Status displays a run time in minutes elapsed from the<br />

start of the run.<br />

The actual cell temperature, the conductance and total conductance as<br />

well as the range will be displayed.<br />

Direct Control<br />

All settings you have entered in this area will become effective after<br />

clicking on the Apply buttons. To achieve changes during a run, the<br />

direct control option must be activated (see Fig. 249 on page 170). Also<br />

an autozero can be initialized.<br />

Instrument Status – User Defined Detector<br />

The detector status tab of a user defined detector displays only the<br />

monitor area. No direct control is possible.