136 Creating an Instrument Control Method<br />

Setting up the Baseline Check<br />

This check is only available if you have activated it according to the next<br />

figure. For configuration of the system, refer to page 38.<br />

Fig. 197 Activation of the baseline check option<br />

Any configured HPLC system with activated baseline check option will be<br />

completed by an additional tab for this check. Click on the baseline<br />

check tab to define the check conditions.<br />

Fig. 198 Defining the baseline check conditions<br />

General<br />

These selections will be valid for all active detectors or detector channels.<br />

Set the time for which the baseline must meet the set criteria to pass the<br />

check.<br />

The check and the method will be stopped, if the system has not passed<br />

the check within the set time. This time must be at least equal to the set<br />

pass time.<br />

Noise Test Method<br />

The noise test will be performed either according to the root of the mean<br />

of squares (rms) method or to ASTM. Select your choice.<br />

For any available detector channel of the configured system a separate<br />

line will appear in the spreadsheet. Here you can set the thresholds for<br />

the noise and drift for each channel separately. Any test (noise and drift)<br />

must be enabled separately for any channel.