164 Creating an Instrument Control Method<br />

Reset Transport Vials (only with 84+3 vials tray)<br />

If the tray type 84+3 vials is configured, the button Reset Transport Vials<br />

allows for reset the transport liquid level of all transport vials (vials 85, 86,<br />

87) to the default value of 8000 µl. If you click on that button, the software<br />

will ask if you wants to reset the filling level of all transport vials to 8000<br />

µl. Click to do so or to leave the window without a level<br />

reset. Please be sure, that you also fill the transport vial(s), because the<br />

autosampler adjusts the needle immersion depth dependent on the vials<br />

filling level. If the vials will not fill, the autosampler can aspirate air instead<br />

transport liquid from the transport vial. The software cannot read-out or<br />

set the filling level. The transport vial filling level will automatically be<br />

reset, if the autosampler will be switched on.<br />

Start Needle Wash<br />

Press this button to force the autosampler to perform the needle wash.<br />

Move Tray<br />

Clicking briefly on this button will change the tray position:<br />

FRONT �� HOME.<br />

[>>…]<br />

The Stacked Injection table can be changed during a run. To have access,<br />

the option Direct Control during a run in Method – Runtime settings must<br />

be enabled. Pressing [>>…]-button during a run will open the known<br />

Stacked Injection setup with the default values. For the re-programming of<br />

the Stacked injection table, please refer the instrument setup chapter of the<br />

autosampler 3950. If you click on the Apply-button, the new stacked<br />

injection table will be executed immediately; the program will not check<br />

when the last injection was made.<br />

If the Direct Control during a run-option Save changes in time table is<br />

enabled, all program lines for stacked injections, which have not been<br />

executed yet during the current run, will be replaced by the stacked<br />

injections from the direct control, whereas lines for stacked injections,<br />

which have already been executing during the current run, remain<br />

untouched.<br />

If the Direct control during a run-option Save changes in time table is<br />

not active, the new table will be executed, but not stored in the instrument<br />

setup. However, in this case the method does not include the correct<br />

injection program.