122 Creating an Instrument Control Method<br />

Fig. 175 Autosampler Triathlon setup window<br />

Standard Injection<br />

In the Standard Injection section you can enter parameters for the<br />

standard actions that can be performed with the autosampler.<br />

Injection Method<br />

Click one of the Flushed, Partial or µl Pick-up fields to select the<br />

injection method.<br />

The maximum programmable injection volumes are:<br />

Flushed loop: not programmable, is equal to the loop volume but<br />

needs more sample to fill the loop completely<br />

(3 x loop volume for loop volumes < 100 µl and 2 x loop<br />

volume for loop volumes > 100 µl)<br />

Partial loop fill: 50 % of the programmed loop volume<br />

µl Pick-up: Injection volume = (loop volume – 3 x needle volume) / 2<br />

Syringe Speed and Scale Factor<br />

The aspirating speed of the syringe used in injection methods can be<br />

adapted depending on the viscosity of the samples (the choices are Low,<br />

Normal, and High). Alternatively, syringe speeds can be reduced by<br />

entering a scale factor (1 to 10). The syringe speed will be the scale<br />

factor multiplied by the syringe speed. The syringe speed during the<br />

washing or the rinsing procedures is not affected by this setting.<br />

Air Segment<br />

Check the Air Segment box to enable the use of an air segment between<br />

the sample and the wash solvent to reduce the amount of flush volume.<br />

Headspace Pressure<br />

Check the Headspace Pressure box to enable the use of headspace<br />

pressure in combination with the syringe to transport sample to the loop.<br />

The compressor will always be used during a wash procedure.<br />

The accuracy and reproducibility of the autosampler may decrease<br />

if headspace pressure is switched off.<br />

Needle Wash<br />

Check the Needle Wash box to enable a wash procedure.<br />

Flush Volume<br />

Enter the amount of sample in µl taken from a vial to preflush the needle<br />

before the loop is filled with the sample. The default value is 30 µl for an<br />

installed tube volume of 15 µl.