38 Setup and Control of Knauer HPLC Systems<br />

A list of available analysis options will be presented.<br />

Fig. 43 analysis options window<br />

Click the box(es) to select the optional software you want to enable for<br />

this instrument. The additional licenses are required beside the Baseline<br />

Check. For more information regarding the option, please check the<br />

System Options part above on page 14. The Knauer Fraction Collectors<br />

Control option (FRC) must not been enabled here. Whenever a<br />

configuration with a fraction collector is started, an installed FRC license<br />

will be activated.<br />

Configuration – Knauer Pumps<br />

For each pump in the Configured Modules window, double-click the icon<br />

and complete the configuration dialog.<br />

For configuration information regarding the PLATINblue pump P-1,<br />

please refer to the PLATINblue system manual.<br />

If the pump S 1050 should be controlled by RS-232, in the pump’s<br />

own setup the “Interface” must be set to “RS232 115200”. Please<br />

refer to the pump’s manual, chapter “Setup menu” for more<br />

information.<br />

The pumps K-1000, K-1001, S 1000 and K-1800 (preparative pump<br />

1800) only can be controlled with ChromGate, if the pump’s internal<br />

communication in the menu option “CONTROL” is set to “NET baud<br />

rate 9600”.<br />

To control a pump K-501, in the pump’s internal setup the menu<br />

option “CONTROL” must be set to “Ext:RS232”. The firmware<br />

version must be v. 1.23 or higher.<br />

If you are operating a K-1001 pump fitted with a firmware version<br />

< 5.x, you must select the K-1000 icon for configuration. This is due<br />

to the status information, which the older firmware does not<br />

support. Otherwise you will get error messages when the pump<br />

starts running. For those pumps no advanced status information<br />

and no pressure trace recording are available.