Setup and Control of Knauer HPLC Systems 67<br />

Fig. 95 Tray configuration for the autosampler 3950 with option Use Mix & Dilute<br />

For µl pick-up no transport vial must be chosen. For the tray type 84+3<br />

vials the transport vial can be selected in the method setup, for all other<br />

tray types the transport liquid will be taken from the bottle with wash<br />

solution.<br />

All configuration settings will be transmitted to the autosampler if<br />

the instrument will be opened. In difference to the other Sparkmanufactured<br />

autosamplers, the configuration data will not be<br />

stored permanently into the device and compared with your<br />

settings. However, please make sure, that the configuration settings<br />

are correct due to that the device will take over your settings<br />

without verification.<br />

When complete, click OK to exit the dialog and return to the instrument<br />

configuration icon list.<br />

Configuration – Autosampler Knauer Optimas<br />

Fig. 96 Configuration window for the autosampler Knauer Optimas<br />

Name<br />

Enter a descriptive name for the sampler. This name will appear on the<br />

autosampler tab when the instrument setup or status windows are<br />

displayed. Use a name which is unique within the instrument.<br />

Serial Number<br />

Type in the serial number of the detector you are using.