68 Setup and Control of Knauer HPLC Systems<br />

Device ID<br />

Select the device identifier for the autosampler from the drop-down list.<br />

This must match the settings of the autosampler being used. To check<br />

the device identifier for the autosampler Knauer Optimas, press the<br />

SYSTEM, MENU and then SERIAL buttons on the autosampler<br />

keyboard. The default identifier is 60.<br />

Serial Port<br />

Select the number for the communication port on your PC where the<br />

autosampler is connected. On the autosamplers rear panel use the port<br />

S2 IN, to connect the autosampler with the computer. It is recommended<br />

to switch all DIP switches for the port to OFF position for a proper<br />

communication.<br />

Loop Volume<br />

Enter the volume of the installed loop in µl. If the Prep mode is enabled,<br />

the loop volume is fixed on 10000 µl.<br />

Tubing Volume<br />

Type in volume of the installed needle and tubing in µl. The valid range is<br />

0 - 999 µl. The volume is noted on a label on the tube.<br />

Syringe Volume<br />

Select the volume of installed syringe from the drop-down list. The<br />

choices are 250 µl, 1000 µl, and 2500 µl. If the 2500 µl syringe is<br />

selected, automatically the Prep mode check box is enabled. In the Prep<br />

mode the values for the syringe (2500 µl) and loop (10000 µl) and the 24<br />

vial tray type are fixed.<br />

Options (Tray cooling / Prep mode)<br />

Check the appropriate boxes if you want to control the tray cooling or<br />

enable the Prep mode. The desired tray temperature can be entered<br />

later as part of the instrument setup. In the Prep mode the values for<br />

syringe (2500 µl) and loop (10000 µl) and the 24 vial tray type are fixed.<br />

The tubing volume is set to 45 µl. Please check the correct volume of the<br />

tube between the needle and the injection valve on the tube’s label.<br />

Tray...<br />

Click the Tray button to set up the autosampler tray: