120 Creating an Instrument Control Method<br />

will be made per run. Please make sure that you have enough samples<br />

for the entered number of injection.<br />

Sample Vial<br />

Select the vials the stacked injection should be made from. Select Same<br />

to injection from the same vial as the previous injection, Next to inject<br />

from the vial number from the previous injection or Custom to<br />

select free the vial number.<br />

Injection Volume (µl)<br />

Select the desired injection volume. Note that the volume depend from<br />

the selected injection mode of the autosampler instrument setup.<br />

Sample Preparation Time (min)<br />

Define the time before the injection time; the autosampler should start<br />

with the sample/injection preparation. Please make sure, that the given<br />

time is sufficient to prepare the injection before the injection time arrives,<br />

otherwise the injection may be delayed. If the autosampler is ready to<br />

inject, it will wait until the programmed time for the stacked injection. If the<br />

autosampler needs a longer time than programmed, it will inject<br />

immediately after the injection preparation is finished. In this case the<br />

next stacked injection will be delayed by this “extra time”.<br />

Time between Injections (min)<br />

Enter the time between the injections.<br />

Do not wash needle between injections from same vial<br />

Enable this option if you wish to prohibit the needle wash after each<br />

injection, programmed in the autosamplers instrument setup, for the<br />

stacked injections from the same vial.<br />

If you close the Stacked Injections window with Apply, the selected<br />

settings will be copied into the Stacked injections table. Any existing<br />

entries from previous programming will be deleted.<br />

If you make a right mouse click on the table, you can select<br />

- Fill Down, to fill the table with the same settings as in the selected line<br />

- Insert Line, to add a new line<br />

- Delete Line, to delete the selected line<br />

- Delete Lines, to delete all selected lines<br />

- Fill Table…, will open the Stacked injection setup again. The new table<br />

will overwrite the existing one. If you select the command Fill table to<br />

open the setup dialog for the stacked injection table, the window will<br />

show default values for all parameters, beside the sample preparation<br />

time.<br />

Click on OK to close the Stacked Injections table window; the<br />

autosampler S 3950 instrument setup window will be shown.<br />

In the Aux Traces tab a trace “Stacked Inj. [AS 3950] can be enabled.<br />

The trace will show the start of the sample preparation and the stacked<br />

injection.