86 Creating an Instrument Control Method<br />

Fig. 124 Error message when pressure limits are out or range<br />

Solvent(s)<br />

Fig. 125 Solvent description in pump setup window<br />

The number of displayed rows depends on the configuration. Isocratic<br />

systems only show the component A. High pressure gradients show up to<br />

four components corresponding to the system configuration. In case of<br />

low pressure gradient pumps three (P2.1L, preparative pump 1800 and<br />

Kontron 42x) or four (S 1050, S 1000, K-1000, K-1001) components can<br />

be defined.<br />

ChromGate ® does not distinguish the two different LPG valve<br />

blocks for the preparative pumps P2.1L and 1800 (refer the pumps<br />

manual). The working limits of these blocks are:<br />

three solvents with a maximum summarized flow of 220 ml/min or<br />

two solvents with a maximum summarized flow of 1000 ml/min. Be<br />

sure to select the correct gradient mode for the connected valves,<br />

otherwise the pump or valves may damaged.<br />

In the K-501 pump’s internal setup screen in the CONTROL menu,<br />

you should set the “Ext:RS232” mode to be able to control the<br />

pump from the ChromGate ® software.<br />

Using the Kontron pump 322 additionally the Compression stage can be<br />

selected. The choices are Low, Medium, and High.<br />

Fig. 126 Solvent description in Kontron 322 pump setup window<br />

Flow OFF at the end of run<br />

Checking this option the flow will be set to zero after elapsing the runtime.<br />

[Action] at the end of run (S 1050 / P2.1L only)<br />

The pumps S 1050 and P2.1L allow not only setting the flow to zero, but<br />

also to switch the pump into Standby after the run.