Creating an Instrument Control Method 111<br />

Fig. 161 ASM 2.1L setup window for fractionation valve<br />

Instrument Setup – Autosamplers<br />

In any system you can include only one autosampler. Therefore all<br />

supported samplers will be explained separately without cross<br />

references.<br />

Instrument Setup – Autosampler 3800<br />

Smartline Autosampler 3800 and K-3800 (Basic Marathon)<br />

The method parameters required for automatic operation of your<br />

autosampler are contained on the Method – Instrument Setup – AS 3800<br />

tab. The control parameters for the autosampler become part of your<br />

method and sequence files.<br />

Fig. 162 Autosampler 3800 setup window<br />

Injection Type<br />

The AS 3800 can perform injections based on the flushed loop or partial<br />

loop filling (for the Plus only) principles. Loop filling can be done on a<br />

volume basis, controlled by the digital dispenser system, or on a time