Setup and Control of Knauer HPLC Systems 55<br />

Configuration – Kontron UV-Detectors<br />

Fig. 79 Kontron detector configuration windows<br />

The Kontron detectors are controlled similar to the Knauer detectors via<br />

the configured RS232 port. For the data acquisition, the analog output<br />

with an A/D card or box must be used.<br />

Name<br />

Enter a descriptive name for the detector. This name will appear on the<br />

detector tab when the instrument setup or status windows are displayed.<br />

The default detector name will be suggested automatically by selecting<br />

the detector type. The name must be unique within the instrument.<br />

Type<br />

Select the type of your detector.<br />

Serial Port<br />

Select the number for the communication port on your PC where the<br />

instrument is connected. The selection is for simulations. No real<br />

control of the instrument will be performed.<br />

Serial Number<br />

Type in the serial number of the detector you are using.<br />

Flow Cell<br />

Enter a description for the flow cell you are using. The description will be<br />

printed out when you print the instrument configuration.<br />

Data acquisition interface<br />

The data acquisition is performed as it was described for the user defined<br />

detectors via the interface box or the Manager 5000/5050. Select the<br />

connected interface from the pull down menu. Most of the Kontron<br />

detectors do not support data acquisition via serial port; therefore the A/D<br />

converter is required.<br />

Channel 1 Input:<br />

Select the connected data channel of your interface.