158 Creating an Instrument Control Method<br />

The right-handed part of the HPG section allows for direct control of the<br />

pumps, independently from the method setup. Please note, that only %A<br />

can be entered, the %B component of the gradient will be completed<br />

automatically by the software. Only a parameter, the check box is<br />

enabled for, can be edited and the changed value will be send by clicking<br />

the -button.<br />

Flow (Direct Control)<br />

To enter a flow, enable the Flow check box. If enabled, you can enter a<br />

flow, which will be the resulting flow of both pumps. Click on the upper<br />

button (same line as for the Flow and A(%) values) to start the<br />

pumps or change the current flow. If the flow rate is out of the acceptable<br />

flow range, defined by the mounted pump head, an appropriate error<br />

message will be shown. If the check boxes for Flow and A(%) have been<br />

enabled, the values for both will be send to the pumps while clicking the<br />

button. If you wants to update only the flow or gradient, please<br />

disable the A(%) check box.<br />

Click on the button to set the flow rate to 0 (zero).<br />

A(%) (Direct Control)<br />

If the corresponding check box is enabled, you can change the gradient<br />

composition. Enter the desired value for HPG pump A and click on the<br />

upper button (same line as for the Flow and A(%) values) to<br />

send the new gradient composition to the pumps. The program will<br />

complete automatically the value for HPG pump B. If the check boxes for<br />

Flow and A(%) have been enabled, the values for both will be send to<br />

the pumps while clicking the button. If you wants to update only<br />

the A(%), please disable the flow check box.<br />

Pmin / Pmax (only for P4.1S) (Direct Control)<br />

Enter the desired values for the allowed minimum and maximum system<br />

pressure.<br />

Click on the button to send the pressure limits to the pump. If the<br />

pressure limits are out of acceptable pressure range, defined by the<br />

mounted pump head, an appropriate error message will be shown.<br />

UV<br />

The left-handed part of the UV section shows the detector status as<br />

Deuterium lamp status, wave length in nm and signal. The signals unit<br />

depends on the unit selected in the detector’s configuration (AU, mAU,<br />

µAU).<br />

Fig. 234 ASM2.1L status tab – UV<br />

D2 Lamp (Status)<br />

The button shows the current lamp status. The On state is represented<br />

by a green active Light Emission Diode LED symbol ( ), the Off state,<br />

including initializing (heating), is represented by an inactive grey LED<br />

symbol ( ).<br />

WL: (Status)