188 ChromGate® System Suitability Setup<br />

Fig. 281<br />

The first sample in your suitability set should be designated as Begin<br />

System Suitability. Additional suitability standards should be designated<br />

as System Suitability Standards, and the final sample in your suitability<br />

set should be designated as End System Suitability. Multiple sample<br />

run types can be selected for a given sample.<br />

A System Suitability Report will be generated at the end of the set of<br />

suitability standards when the sequence is run.<br />

To view the Suitability Report on screen, click Reports/View/Sequence<br />

Custom Report choice. The System Suitability Report will appear in a list<br />

of available Sequence Custom Reports.<br />

To print the Suitability Report, click on Reports/Print/Sequence Custom<br />

Report, and select the System Suitability Report choice. Note that in<br />

order to have your report printed automatically at the end of the<br />

sequence, you must have selected the Print Sequence Reports option<br />

in the Sequence Properties dialog.<br />

Suitability Reports<br />

EZChrom Elite Client/Server provides a default suitability report template.<br />

This report template can be used as-is, or it can be customized in the<br />

Sequence Custom Report editor.