Creating an Instrument Control Method 135<br />

Setting up Auxiliary Traces<br />

Fig. 195 Setup window, Auxiliary Traces Any configured HPLC system with at<br />

least one pump with pressure transducer, a column oven or a flowmeter<br />

will be completed by an additional tab for auxiliary traces.<br />

Only the selected traces will be stored with the run data.<br />

Setting up a Trigger<br />

Any configured HPLC system with at least one detector acquisition<br />

channel will be completed by an additional tab for the trigger.<br />

Click on the Trigger tab to designate the trigger type and to setup the<br />

synchronization. The trigger type determines how the data sampling<br />

and the gradient program(s) are started.<br />

Fig. 196 Setup window, trigger tab<br />

Trigger Type<br />

Select the type of synchronization from the drop-down list. The choices<br />

are None, <strong>Manual</strong>, and External.<br />

None Time program(s) and Sampling start immediately after clicking<br />

on Start. Sequence acquisitions do not pause between runs.<br />

<strong>Manual</strong> Operator must press Enter to start the run. Sequence<br />

acquisitions pause for confirmation between runs.<br />

External Program execution and data sampling are started from an<br />

external trigger, if this option is selected. Further settings are<br />

not necessary. Sequence acquisitions pause for external<br />

trigger signal between runs.