Creating an Instrument Control Method 149<br />

Autozero<br />

Click this button to define the current signal as a zero level.<br />

Scan, nm<br />

This section is disabled for changes. It only provides information on how<br />

the scan spectral interval is specified. Changes must be made from the<br />

instrument setup window (Fig. 150 on page 102). It is not possible to<br />

modify this function during a run.<br />

Digital Channels (nm)<br />

The number of channels that will be activated depends on how many<br />

possibilities are given to set new values for the measuring wavelengths.<br />

Pressing the WL Apply button forces the wavelength change on the<br />

detector channels.<br />

Digital Outs (Events)<br />

Click the button to set up the digital outputs. Refer to the chapter<br />

Instrument Setup – Pumps, Pump section Configuring Digital Outputs for<br />

details. Click the DO Apply button to accept the settings. For detectors<br />

without the feature of digital outputs this area is grayed and inaccessible.<br />

The D2 lamp should be switched on manually from the direct control<br />

window or using the Wake-Up option from the K-2600 before<br />

starting the next working session. After power up, allow 30 minutes<br />

for the lamp to stabilize before running samples that may require<br />

sensitive measurements.<br />

Instrument Status – Diode Array Detectors<br />

(S 2600, DAD 2850, DAD 2800, K-2700 and Kontron DAD 540/545)<br />

Fig. 223 DAD-Detector status tab, example: DAD 2600<br />

Monitor<br />

Run Status<br />

The status HOLD will be displayed when no program is running;<br />

otherwise the Run Status displays a run time in minutes elapsed from the<br />

start of the run.<br />

Channel 1 / 2 / 3 / 4<br />

The actual signal in the units which have been chosen in the detector<br />

configuration dialog at the actual wavelength in nm is displayed for each<br />

active channel.<br />

Direct Control<br />

The direct control option must be activated (see Fig. 249 on page 170).