90 Creating an Instrument Control Method<br />

Pretreatment (not for Kontron pumps)<br />

Checking this option the Setup… button becomes active. Clicking on it<br />

the following dialog window will be opened:<br />

Fig. 133 Pretreatment setting dialog window<br />

The appearance of the pump program corresponds to that of the main<br />

pump setup window. It must be edited in the same way as described<br />

above. To add a new line, make a right mouse click into the table and<br />

select “Insert line” from the menu. The only difference is that negative<br />

times must be entered, because they are relative to the zero injection<br />

time.<br />

During a pretreatment, no data acquisition is possible. However, the run<br />

status during a pretreatment is “Equilibrating”.<br />

If a pretreatment is to define for more than one instrument (pump<br />

and/or valve drive) in a method, it is required to set for all of them<br />

the same earliest time. All pretreatment procedures will start<br />

simultaneously.<br />

If a pump should exactly switch off after 5 minutes of pretreatment,<br />

include a line with 0 ml/min flow into the pretreatment’s pump<br />

program. Otherwise the pump may run for a longer time as it need<br />

some seconds to load the new flow rate from the method’s “normal”<br />

pump program<br />

An example for both :<br />

A pump should run for exactly 5 minutes before the injection and a valve<br />

should switch 2 minutes before injection from position 2 to position 5,<br />

means, 3 minutes after the pump’s pretreatment is started. Here are the<br />

setup tables create a time table for the valve, which starts with -5<br />

minutes:<br />

Pump<br />

-5.04 min 1 ml/min<br />

-0.04 min 1 ml/min<br />

-0.02 min 0 ml/min (smallest time step is 0.06 minutes for pump 1800 /<br />

P2.1L or 0.02 min for all other pumps)<br />

0.00 min 0 ml/min<br />

Valve<br />

-5.04 min pos. 2<br />

-2.00 min pos. 5<br />

0.00 min pos. 5