52 Setup and Control of Knauer HPLC Systems<br />

Before you open the ChromGate instrument you must switch the<br />

detector in to serial mode.<br />

RF-10 Axl: Press the function key on the detector’s keypad<br />

as often as "RS232C" will be displayed. Confirm with "E" key. The<br />

display will show "CONNECT", also confirm with "E" key.<br />

If you press the "function" while "CONNECT" will be displayed, you<br />

will enter to the COM connection table. Changed values must be<br />

confirmed by pressing the key “E”, to move to the next parameter,<br />

press the function key.<br />

Parameter Value<br />

Baud 6 (means 9600)<br />

Data Bits 8<br />

Parity 2 (means even parity)<br />

Stop Bit 1<br />

The COM settings here and in the software must be identical. If,<br />

e.g., the parity is set to “0”, means no parity, it must be set in<br />

the detector’s setup as well in the software.<br />

If all parameters have been entered, the display shows the message<br />

“CLOSE KEY”. Press key “E” to leave the setup. To switch the<br />

device to serial control mode, please follow the steps above<br />

To leave the serial mode, press the shift and CE/del key at the same<br />

time.<br />

RF-20A/Axs: Press the function key until the display shows<br />

“SYSTEM _ Enter to Select”. Press the key. Press function<br />

key until the display shows “RS-232 PARAM _ Enter to select”.<br />

Press the key. Now the settings for the serial interface can<br />

be reviewed and modified if necessary. To move to the next<br />

parameter, press the function key<br />

Parameter Value<br />

Baud 9600<br />

Data Bits 8<br />

Parity 0 (means no parity)<br />

Stop Bit 1<br />

The COM settings here and in the software must be identical. If,<br />

e.g., the parity is set to “2”, means even parity, it must be set in<br />

the detector’s setup as well in the software.<br />

If all parameters have been reviewed, the display shows “RS-232C _<br />

Enter to Enable”. Press the key . The display shows “RS-<br />

232C _ Enable” for only about 2 seconds, then the default screen<br />

will be shown. Now the detector is in serial control mode and can be<br />

controlled by software.<br />

If the detector is switched Off and On, it will start again in RS-232<br />

control mode. During the boot procedure the RS-232 control mode<br />

will be shown very shortly on the display.<br />

To switch off the RS-232 control mode, the keys und <br />

must be pressed at the same time.