52 Setup and Control of Knauer HPLC Systems<br />
Before you open the ChromGate instrument you must switch the<br />
detector in to serial mode.<br />
RF-10 Axl: Press the function key on the detector’s keypad<br />
as often as "RS232C" will be displayed. Confirm with "E" key. The<br />
display will show "CONNECT", also confirm with "E" key.<br />
If you press the "function" while "CONNECT" will be displayed, you<br />
will enter to the COM connection table. Changed values must be<br />
confirmed by pressing the key “E”, to move to the next parameter,<br />
press the function key.<br />
Parameter Value<br />
Baud 6 (means 9600)<br />
Data Bits 8<br />
Parity 2 (means even parity)<br />
Stop Bit 1<br />
The COM settings here and in the software must be identical. If,<br />
e.g., the parity is set to “0”, means no parity, it must be set in<br />
the detector’s setup as well in the software.<br />
If all parameters have been entered, the display shows the message<br />
“CLOSE KEY”. Press key “E” to leave the setup. To switch the<br />
device to serial control mode, please follow the steps above<br />
To leave the serial mode, press the shift and CE/del key at the same<br />
time.<br />
RF-20A/Axs: Press the function key until the display shows<br />
“SYSTEM _ Enter to Select”. Press the key. Press function<br />
key until the display shows “RS-232 PARAM _ Enter to select”.<br />
Press the key. Now the settings for the serial interface can<br />
be reviewed and modified if necessary. To move to the next<br />
parameter, press the function key<br />
Parameter Value<br />
Baud 9600<br />
Data Bits 8<br />
Parity 0 (means no parity)<br />
Stop Bit 1<br />
The COM settings here and in the software must be identical. If,<br />
e.g., the parity is set to “2”, means even parity, it must be set in<br />
the detector’s setup as well in the software.<br />
If all parameters have been reviewed, the display shows “RS-232C _<br />
Enter to Enable”. Press the key . The display shows “RS-<br />
232C _ Enable” for only about 2 seconds, then the default screen<br />
will be shown. Now the detector is in serial control mode and can be<br />
controlled by software.<br />
If the detector is switched Off and On, it will start again in RS-232<br />
control mode. During the boot procedure the RS-232 control mode<br />
will be shown very shortly on the display.<br />
To switch off the RS-232 control mode, the keys und <br />
must be pressed at the same time.