118 Creating an Instrument Control Method<br />

Headspace Pressure<br />

Check the Headspace Pressure box to enable the use of headspace<br />

pressure in combination with the syringe to transport sample to the loop.<br />

The compressor will always be used during a wash procedure.<br />

The accuracy and reproducibility of the autosampler may decrease<br />

if headspace pressure is switched off.<br />

Needle Wash<br />

Check the Needle Wash box to enable a wash procedure.<br />

Temperature control<br />

If your autosampler 3950 is equipped with a tray cooling facility, this<br />

option can be checked and the desired temperature (4 – 22 °C) can be<br />

entered.<br />

Wait with tolerance (±°C)<br />

The system becomes ready only if the actual tray temperature is inside<br />

the set range.<br />

Wait time (min)<br />

The waiting time is the equilibrating time in minutes after the tray temperature<br />

reached the temperature set range before the injection takes<br />

place.<br />

Mix Methods...<br />

Click the Mix Methods button to create a program of sample preparation.<br />

The Mix Methods set up dialog will appear. It is the same dialog as<br />

already described for the AS 3900, starting with Fig. 165 on page 115.<br />

Please note, that for the autosampler 3950 for all tray types except the<br />

83+3 vials tray no transport vial for the µl pick-up can be defined. The<br />

transport liquid will be escaped from the wash solution. If the 84+3 vials<br />

tray is selected, the vials 85, 86 or 87 can be selected as a transport vial.<br />

The transport plug is 2.5 x needle volume (needle volume = tubing<br />

volume, please refer to the autosampler configuration chapter). The<br />

autosampler itself calculates the liquid level in the transport vials. It will be<br />

reset to 8000 µl every time if the autosampler is switched off and on.<br />

Additionally the level can be reset from the autosamplers direct control<br />

tab. The level cannot be read-out from the software. If the autosampler<br />

calculates, that the vial is empty, the method/sequence will stopped.<br />

Dependent from the transport liquid level, the autosampler will adapt the<br />

needle immersion depth. Therefore it is important to fill the transport vials<br />

whenever the autosampler was switch off or the transport liquid level was<br />

reset manually. If a vial is selected as the transport vial, the configuration<br />

of the vial as a Reagent A or B vial will be ignored.<br />

Stacked injections...<br />

The option Stacked injections allow to define additional injections during a<br />

run. Such injections can only be programmed either for the autosampler<br />

3950 or the injection module (please refer the FRC section, chapter<br />

“Stacked Injections” of this manual). Click on the Stacked injections –<br />

button to open the Stacked Injections table dialog.<br />

Enable the check box “Use stacked injection”.