Creating an Instrument Control Method 87<br />

Fig. 127 Possible action at the end of run for pump S 1050 / P2.1L<br />

Pretreatment<br />

This option enables several pretreatment operations of pumps and valves<br />

before injection and starting the data acquisition. For setting up see<br />

below on page 89.<br />

Solvent Type… (S 1050 / P2.1L only)<br />

Select in the Solvent Type menu the used solvent for a proper<br />

compression and flow correction. Note that for an LPG pump only one<br />

solvent type can be selected.<br />

Fig. 128 Solvent type settings for pump S 1050 / P2.1L<br />

Pump Program<br />

Accordingly the spreadsheet for the pump program is displayed. Up to<br />

four columns for the composition will be displayed. Additionally, for<br />

pumps with digital outputs, the option for digital outs is given in a<br />

separate column. The spreadsheets of the S 100 / 10P/20P/P2.1S/P4.1S,<br />

K-120, K-501, and the Kontron pumps do not include this column.<br />

A<br />

B<br />

Fig. 129 Pump program in pump setup window: A Pump 1800 LPG, B K-501<br />

isocratic<br />

For all time programs, the minimum time step between two lines is<br />

0.02 min, except for the preparative pumps P2.1L / 1800; here the<br />

minimum time step is 0.06 min. For the pump P2.1L in isobar mode,<br />

the smallest time step is 0.25 minutes.<br />

Use the spreadsheet in the pump setup view to enter your time (gradient)<br />

program. The number of columns shown depends on the defined system<br />

(isocratic, LPG, or HPG) and the number of included components.