Creating an Instrument Control Method 147<br />

Instrument Status – Kontron Detectors (3xx, 4xx, and 5xx)<br />

Fig. 220 Detector status tab, Kontron 430<br />

Monitor<br />

Run Status<br />

The status No run will be displayed when no program is running;<br />

otherwise the Run Status displays a run time in minutes elapsed from the<br />

start of the run.<br />

Channel 1 / 2<br />

The actual signal in the units which have been chosen in the detector<br />

configuration dialog at the actual wavelength in nm is displayed for each<br />

active channel.<br />

Communication – instrument<br />

This field displays the serial port of the detector which is used to control<br />

the instrument. The active state of the communication with an instrument<br />

is represented by the green symbol. If the serial port was selected as<br />

, then no real control of the detector takes place.<br />

Communication - A/D interface<br />

The active state of the communication with an A/D interface is<br />

represented by the green symbol. Only in this case you will have<br />

access to the Device Information. Click the button to display the<br />

instrument(s) connected to that COM port. Information about the serial<br />

number, device name, and instrument software versions will appear:<br />

Fig. 221 Information about the connected instrument(s)<br />

Direct Control<br />

Acquisition Channels<br />

The number of channels that will be activated depends on how many<br />

possibilities are given to set new values for the measuring wavelengths.<br />

Pressing the Apply button forces the wavelength change on the detector<br />

channels.<br />

Acquisition Parameters<br />

Select the Resp. time from the drop-down list. Smaller Resp. time (or<br />

time constants) are required for high speed analyses. A good rule is to