Creating an Instrument Control Method 93<br />

Instrument Setup – UV Detectors (UVD 2.1L, UVD 2.1S, S 2520,<br />

S 2500, K-2001, K-2501)<br />

For the detectors K-2001/2501 you should set in the detector setup<br />

screen / EXT.CONTROL menu the keyboard mode to be able to<br />

control the wavelength from the ChromGate ® software.<br />

The detector basic settings, as time constant, sampling rate etc. are<br />

described on page 91.<br />

Fig. 136 UV Detector setup window<br />

Autozero at start<br />

Check the Autozero field if you wish the signal to be zeroed at the<br />

moment a measurement is started (to define the current signal as the<br />

baseline).<br />

Autozero at WL change<br />

No access to this option. The grayed area is only a reminder, that the<br />

autozero will take place at any wavelength change during a run.<br />

Lamp OFF at the end of run<br />

Check the lamp OFF box to switch off the detector lamp after a<br />

measurement is finished. This option may be used for the shutdown<br />

methods.<br />

Detectors UVD 2.1L, S 2520 and 2550: If the lamp was switched off<br />

by the method and you switch off the detector then, on the next<br />

detector start the lamp is still off; it must be switched on manually<br />

in the detector’s own setup or with the software via Instrument<br />

Status – Direct Control.<br />

[Action] at the end of run (UVD 2.1L, UVD 2.1S, S2520 only)<br />

The detectors S UVD 2.1L, UVD 2.1S and S 2520 allow not only to switch<br />

off the lamp at the end of run, but also to switch the device into Standby.<br />

You should switch the lamp on manually before you start the next<br />

working session. After power up, allow 30 minutes for the lamp to<br />

stabilize before running samples that may require sensitive<br />

measurements.