152 Creating an Instrument Control Method<br />

results in an increased signal to noise ratio because the multiple<br />

measurements cause the random noise to be cancelled out. That is, if the<br />

noise is truly random, more measurements will decrease the noise<br />

because for each positive excursion in the noise will have a good<br />

likelihood of being cancelled by a negative excursion. But the signal,<br />

since it is not random, will remain the same over multiple measurements.<br />

Bandwidth is a tradeoff, with a high value decreasing the noise but also<br />

decreasing the possible resolution which can be measured. 5 nm is the<br />

default value, but if your peaks have substantially wider UV spectra you<br />

can increase the bandwidth value in ChromGate ® to obtain a lower noise<br />

level.<br />

Mode<br />

Select the signal mode for your measurements from the drop-down list.<br />

The choices are Absorption and Intensity.<br />

Open/Close Shutter<br />

A click on this button will open/close the shutter. The label will be<br />

changed accordingly.<br />

Filter On/Off (S 2600 only)<br />

You can click on this button depending on whether or not you are using<br />

an edge filter.<br />

Dark Signal<br />

Click the button to measure the dark current. This operation is enabled<br />

only for devices equipped with a shutter.<br />

Set Reference<br />

Click the button to measure a spectrum to be used as a reference<br />

spectrum.<br />

Spectrum<br />

Click the button to measure an absorption or intensity spectrum,<br />

depending on the signal mode selected for the measurement. The<br />

measured spectrum will be shown in the spectrum view.<br />

Instrument Status – Fluorescence Detector RF-10Axl / RF-20A<br />

Fig. 226 Detector Status tab, Fluorescence Detector RF-10Axl<br />

Status<br />

The status HOLD will be displayed when no program is running;<br />

otherwise the Run Status displays a run time in minutes elapsed from the<br />

start of the run.<br />

The actual excitation wavelength, emission wavelength, signal value,<br />

sensitivity, gain, and the selected output mode will be displayed.