250 ChromGate® Preparative Option<br />

To change the settings for colors, font type etc., open the<br />

chromatogram’s appearance (right mouse click on the chromatogram –<br />

Appearance). Select the correct detector channel in Trace and select the<br />

annotation you want to change the settings for.<br />

Fig. 358 Chromatogram - FRC Appearance<br />

Please note, that for report previews as well as for pdf prints, the fraction<br />

color will not be shown correctly (always pink/yellow). If the report is been<br />

printed with a “physical” printer, the colors are as selected in the<br />

Appearances menu.<br />

Stacked Injection<br />

The option Stacked Injection allows to perform injection during a run. As<br />

an injection device the autosampler 3950 or the Injection Module can be<br />

used. For the configuration and instrument setup of the autosampler<br />

3950, please refer the appropriate chapters.<br />

Injection module configuration<br />

The injection module will be installed with the Knauer FRC AddOn. In the<br />

instrument configuration dialog, select the injection module and add it to<br />

the Configured modules. Note that it is not allowed to add an autosampler<br />

and the injection module.<br />

Fig. 359 Injection module configuration window<br />

The injection module consists of a pump and a Knauer 2-position<br />

switching valve. Optional an injection loop can be used. As a pump a<br />

pump P 2.1L, P 2.1S, P 4.1S, S 100, S 1050 or 1800 can be selected. If