172 Knauer Instrument Control Method Options<br />

Fig. 252 <strong>Advanced</strong> Runtime Settings – System wake-up… - date setting<br />

If enabled, this option is valid for the currently used computer.<br />

Trace name<br />

If the option “Trace name” is enabled, the wave length of the initial line of<br />

the wave length table becomes part of the trace name in the<br />

chromatogram window as well as in the report (chromatogram, channel<br />

name in run report) instead of the channel name. This belongs to all<br />

detectors with a wave length table, as UV, DAD and fluorescence<br />

detectors. The channel name in the channel selector will not be changed.<br />

Fig. 253 <strong>Advanced</strong> Runtime Settings – Trace name changes<br />

The option will only change the trace name for new chromatograms.<br />

If enabled, this option is valid for all instruments of the Enterprise; it<br />

is not limited on the currently used instrument or computer. That<br />

means, that the trace name will include the wave length for all<br />

instruments and therefore for all methods that will be stored on the<br />

mentioned Enterprise machine.<br />

Download Tab / Method<br />

The actually set parameters (may be even not saved) of the current<br />

instrument setup tab as well as those of the whole method can be<br />

downloaded to the corresponding instruments. Select the menu<br />

CONTROL – DOWNLOAD METHOD or<br />

CONTROL – DOWNLOAD TAB<br />

Trace name not changed<br />

Trace name changed