Creating an Instrument Control Method 89<br />

Configuring Events / Digital Outputs<br />

This option is given only for the S 1050, P2.1L, S 1000, K-1001, and K-<br />

1800 pumps. The spreadsheets of the S 100 / 10P/20P/P2.1S/P4.1S, K-<br />

120, K-501, and the Kontron pumps do not include this column. Configure<br />

the events by pressing the button which is displayed when you place<br />

the cursor on the Events cell. The following dialog will appear.<br />

Fig. 131 Configuring Digital Outputs<br />

The Events (Digital Outputs) correspond to the outputs available on the<br />

rear panel of the pump.<br />

Select the state for each Event you want to set from the drop-down list.<br />

The choices are Off (0), On (1), and Pulse (P). Refer to the pump’s user<br />

manual for details about digital outputs parameters.<br />

Use the All On or All Off buttons to switch all of the outputs either on or<br />

off.<br />

Constant pressure mode (Isobar) (P2.1L only)<br />

If in the instrument configuration the gradient mode for the pump P2.1L is<br />

set to Isobar, the setup windows looks different:<br />

Fig. 132 Instrument setup window P2.1L constant pressure mode (isobar)<br />

Constant Pressure Mode: Flow Limits (ml/min): Set the allowed flow<br />

range for this operating mode. The allowed absolute minimum is 0<br />

ml/min, the maximum depend on the selected pump head.<br />

Pump program – Pressure (unit): Enter the desired system pressure.<br />

The pump will vary the flow, if the pressure in the system will change.<br />

The pressure unit depends on the unit selected in the pump’s instrument<br />

configuration. The minimum time step between two time lines is 0.25 min.