Setup and Control of Knauer HPLC Systems 57<br />

Halogen Lamp<br />

Check this box when the detector is equipped with a halogen lamp.<br />

Configuration – Virtual Detector<br />

To configure a virtual detector you have only to open and to close the<br />

configuration window without any additional entering. The Y-axis<br />

multiplier is only to change if very weak signals are to simulate.<br />

Fig. 81 Virtual detector configuration window<br />

To avoid possible signal distortion, the Y-Axis multiplier in the<br />

configuration of the virtual detector must be equal to that of the<br />

selected data file.<br />

Detector Connections<br />

Up to four detectors can be connected to the Interface Box or Manager<br />

5000/5050/IF2. The detector connections are principally identical. The<br />

following description for the Interface Box can also be used for the<br />

Manager 5000/5050/IF2, only the connector positions may differ.<br />

Always use a shielded twisted pair cable to connect each detector to the<br />

Interface Box. This minimizes electrical noise. Connect the analog output<br />

lines from each detector you want to acquire data from to one of the four<br />

analog connections labeled INTEGRATOR INPUTS 1 through 4.<br />

After power up, allow 15 minutes for the electronics to stabilize<br />

before running samples that require sensitive measurements.<br />

To an integrator<br />

output of a detector<br />

Trigger Autozero<br />

channel 1<br />

channel 2<br />

channel 3<br />

channel 4<br />

Fig. 82 Cable connections to the Knauer Interface Box, Model 96