134 Creating an Instrument Control Method<br />

Fig. 193 Setup window, column oven Jetstream<br />

For details on the spreadsheet operations, refer to the section Creating a<br />

Time (Gradient) Program as described in the chapter Instrument Setup –<br />

Pumps, Pump Program.<br />

The oven program profile will be adapted automatically to the<br />

spreadsheet inputs (compare Fig. 192).<br />

Wait before injection<br />

If you are working with a temperature program check this option. The<br />

setting Wait with tolerance (+/- °C) defines the accuracy of the oven<br />

temperature allowing a new injection.<br />

A temperature independent delay between succeeding injections can be<br />

set in the field Wait time (min). The system will wait for that time after<br />

having reached the initial temperature and before the next injection is<br />

performed.<br />

Instrument Setup – Flowmeter<br />

The setup of the flowmeters is the same for all supported models is very<br />

simple.<br />

Fig. 194 Setup window, Flowmeter<br />

Run time<br />

Enter the desired run time.<br />

Acquisition delay<br />

Here you can define a delay time between starting the run and the data<br />

acquisition.