62 Setup and Control of Knauer HPLC Systems<br />

Fig. 89 Configuration windows left device tab, device type selection<br />

Device Type<br />

Select the device type that is installed on the left side in the assistant<br />

housing, from the drop-down menu.<br />

Pump P2.1S<br />

Pump P4.1S<br />

If selected, you can choose the device name (please refer to<br />

the device name section below), the gradient type (none, HPG<br />

A, HPG B) and the pump head (10 ml, 50 ml).<br />

If selected, you can choose the device name (please refer to<br />

the device name section below), the gradient type (none, HPG<br />

A, HPG B) and the pump head (10 ml, 50 ml). Due to that the<br />

pump P4.1S is equipped with a pressure sensor, you can also<br />

select the unit for the pressure monitoring (MPa, bar, psi).<br />

Detector UVD2.1S<br />

Knauer Valve<br />

Valco Valve<br />

If selected, you can choose the device name (please refer to<br />

the device name section below) and the Y-Axis Units (µAU,<br />

mAU, AU). The “Flow Cell” field allows for entering information<br />

regarding the used detector flow cell.<br />

If selected, you can choose the device name (please refer to<br />

the device name section below) and the number of position.<br />

Please be sure to select the correct number of position, if you<br />

select it manually. 2, 6, 12 or 16 position can be selected. For<br />

one valve in the assistant you can check the “Use as a<br />

Fraction Collector” box. If selected, there are additional<br />

options: You can enter the Vial Volume (please be sure to<br />

enter the correct volume) and setup Tubing and Solvent/Peak<br />

Recycling. Please refer to the chapter Multi Valve Fraction<br />

Collector Configuration for detailed information. The Number<br />

of Vials is fixed and inaccessible, depending on the selected<br />

number of positions of the used valve and the Solvent/Peak<br />

Recycling setup. If for another valve in the assistant the option<br />

“Use as a Fraction Collector” is already selected, for the next<br />

valve an error message “The fraction collection option is<br />

already selected for another valve” will be shown.