62 Setup and Control of Knauer HPLC Systems<br />
Fig. 89 Configuration windows left device tab, device type selection<br />
Device Type<br />
Select the device type that is installed on the left side in the assistant<br />
housing, from the drop-down menu.<br />
Pump P2.1S<br />
Pump P4.1S<br />
If selected, you can choose the device name (please refer to<br />
the device name section below), the gradient type (none, HPG<br />
A, HPG B) and the pump head (10 ml, 50 ml).<br />
If selected, you can choose the device name (please refer to<br />
the device name section below), the gradient type (none, HPG<br />
A, HPG B) and the pump head (10 ml, 50 ml). Due to that the<br />
pump P4.1S is equipped with a pressure sensor, you can also<br />
select the unit for the pressure monitoring (MPa, bar, psi).<br />
Detector UVD2.1S<br />
Knauer Valve<br />
Valco Valve<br />
If selected, you can choose the device name (please refer to<br />
the device name section below) and the Y-Axis Units (µAU,<br />
mAU, AU). The “Flow Cell” field allows for entering information<br />
regarding the used detector flow cell.<br />
If selected, you can choose the device name (please refer to<br />
the device name section below) and the number of position.<br />
Please be sure to select the correct number of position, if you<br />
select it manually. 2, 6, 12 or 16 position can be selected. For<br />
one valve in the assistant you can check the “Use as a<br />
Fraction Collector” box. If selected, there are additional<br />
options: You can enter the Vial Volume (please be sure to<br />
enter the correct volume) and setup Tubing and Solvent/Peak<br />
Recycling. Please refer to the chapter Multi Valve Fraction<br />
Collector Configuration for detailed information. The Number<br />
of Vials is fixed and inaccessible, depending on the selected<br />
number of positions of the used valve and the Solvent/Peak<br />
Recycling setup. If for another valve in the assistant the option<br />
“Use as a Fraction Collector” is already selected, for the next<br />
valve an error message “The fraction collection option is<br />
already selected for another valve” will be shown.