Creating an Instrument Control Method 123<br />

Flush volumes smaller than twice the volume of the needle and<br />

tubing will result in a decrease of performance of the autosampler.<br />

Wash Volume<br />

Enter the volume of wash solvent in µl to be used. The minimum<br />

programmable volume is 300 µl.<br />

Needle Height<br />

Enter the distance between the needlepoint and the plate holder in mm<br />

(default value is 5 mm).<br />

Temperature control<br />

If your Triathlon is equipped with a tray cooling facility, this option can be<br />

checked and the desired temperature can be entered.<br />

Alarm Buzzer<br />

Check the alarm buzzer box to turn ON the alarm beep, given when an<br />

error occurs during a run.<br />

Run Screens<br />

Check the run screens box to display the actual run information during a<br />

run.<br />

Timed events…<br />

These type of events or methods enable you to control optional ISS<br />

valves and other devices via auxiliary (contact closures) or a binary<br />

output (BCD or HEX).<br />

Click the Timed events... button to set up the time base methods. The<br />

time base methods setup dialog will appear.<br />

Fig. 176 Autosampler Triathlon Time base Methods<br />

Click the New button to add one more line to the programmed<br />

methods.<br />

Select the Method Type from the drop-down list. The choices are<br />

AUX-1, AUX-2, AUX-3, AUX-4, ISS-A, ISS-B, CODE-Out, and SSV<br />

Port.<br />

Every auxiliary AUX-1 through AUX-4 can be set to ON (contact<br />

closure) four times and four times to OFF at a programmed Time. It is<br />

not possible to program the same auxiliary to ON and OFF at the<br />

same time.<br />

The ISS valve can only be programmed if the optional ISS valves are<br />

installed. The valves ISS-A and ISS-B can be set to positions 6-1 four<br />

times and four times to 1-2 at a programmed Time.<br />

The SSV port can only be programmed if the SSV option is installed.<br />

Set the time and the SSV port number, a value between 1 and 6.