ChromGate® Preparative Option 255<br />

Fig. 367 Injection Module status window (initial injection)<br />

Monitor<br />

Run Status<br />

Shows the current run status of the module. If no run is performed, the<br />

status is Hold. During the initial injection the status Injecting with the time<br />

of the current injection duration is shown. During the run the run status<br />

shows the run time.<br />

Pump Flow<br />

Shows the current flow rate of the injection module’s pump.<br />

Pump Pressure<br />

Shows the current pressure of the injection module’s pump. This option is<br />

only shown if the pump is equipped with a pressure transducer.<br />

Vial Position<br />

The vial position can be either L (Load) or I (Inject).<br />

If no sample loop is used, that means, that the sample will be injected<br />

directly on the column, the vial position during the injection (while the<br />

injection module’s pump is running) is I (Inject). If the injection is made,<br />

the valve will switch into L (Load) position.<br />

If a sample loop is used, the valve will switch to L (Load) position if the<br />

pump is running and fill the sample loop. For the injection, the valve will<br />

switch to I (Inject) position.<br />

Direct Control<br />

Pump Flow<br />

During a run, the pump cannot be controlled using the pump flow option.<br />

The buttons Stop and Flow Apply will be inaccessible.<br />

[>>…]<br />

If the option Direct control during a run is enabled, after the initial<br />

injection the [>>…]-button is accessible. If you click on the button, the<br />

known stacked injection setup with the default values will open. This is<br />

allow for create a modified stacked injection table. The setup works as<br />

described for the instrument setup of the injection module.