ChromGate® Preparative Option 237<br />

Adjust FRC table after stacked injection<br />

This option allows you to copy the already executed part of the fraction<br />

table for the next injection, if a stacked injection is performed. In this case<br />

you will have the same collection conditions for every injection without<br />

any manual intervention. This makes sense, if you use Single Event<br />

parameters, which otherwise should have to be programmed separately<br />

for every injection. If you use the peak recognition mode, just program the<br />

collection condition for the whole run time, the Adjust FRC table option<br />

leave unchecked.<br />

Pump<br />

If more than one pump is configured in the system, select the used one<br />

from the drop down list for volume and delay calculations.<br />

Update Flow<br />

The set flow of the selected pump after the last actualization will be<br />

displayed. If the flow is not constant for the whole method, the value of<br />

0.000 will be shown. Please enter then the flow rate for the time the<br />

fractions will be collected. The flow is important to calculate the time<br />

delay between detecting a component in the detector and the approach<br />

to the fraction valve as well as the filling level of the vials.<br />

Fraction Collection Program<br />

The editing principles are again the same as it was described for the<br />

pump gradient programming. Some fields are free for entering data<br />

whereas others are either only informative without access or you can<br />

select the possible options via a drop down list.<br />

Exp. Time (min)<br />

Enter the desired time values.<br />

Mode<br />

Fig. 330 Fractionation mode selection<br />

The single event option is to select for programming a detailed exact<br />

fractionation program, whereas the peak recognition is preferably for<br />

use in method development. Every new detected peak according to<br />

the defined thresholds (see Parameters on page 238) will be collected<br />

in the next empty vial. The selected mode will influence the further<br />

setting possibilities.<br />

Event<br />

Fig. 331 Event definition<br />

The possible options depend on the selected mode: