SEC Option 271<br />

Std Mark-Houwink k<br />

If you have selected Mark-Houwink for the Use Constants field, enter the<br />

k value for the standard in this field.<br />

Std Mark-Houwink a<br />

If you have selected Mark-Houwink for the Use Constants field, enter the<br />

a value for the standard in this field.<br />

Intrinsic Viscosity<br />

If you have selected Intrinsic Viscosity for the Use Constants field, enter<br />

the value here. Otherwise this field is not active.<br />

Time Slice<br />

Select a time slice width (in seconds) for the calculations. The slice width<br />

will determine the frequency of Molecular Mass fractions reported. If a<br />

Slice Report is generated, it will be based on the slice width you select<br />

here. The software will force this number to be an integral of the data<br />

collection rate during the analysis.<br />

Baseline Start and Stop Times<br />

The baseline start and stop times (in minutes) can be entered manually<br />

or selected graphically using the SEC buttons on the chromatogram<br />

toolbar.<br />

Results Range Start and Stop<br />

One or more ranges of the chromatogram must be defined to be used in<br />

calculation and reporting of the Molecular Mass distribution. Each range<br />

is determined by the results range start and stop values. The times (in<br />

minutes) can be entered manually or graphically using a stored<br />

chromatogram.<br />

Horizontal Baseline<br />

Select this button if you wish the SEC baseline to be drawn as a<br />

horizontal line forward from the Baseline Start Time to the Stop Time.<br />

The horizontal baseline stops when it crosses the data trace. A threshold<br />

value in microvolt must be specified to ignore any detector baseline noise<br />

in the trace that would cause the horizontal projection to stop when it<br />

encounters the noise in the trace. If the Horizontal Baseline reaches the<br />

end of the detected peak before the programmed Stop Time, the range<br />

will stop where the Horizontal Baseline crosses the data trace.<br />

Calibration<br />

Select Universal for the calibration type.<br />

If Universal Calibration is selected, the Mark-Houwink constants or<br />

intrinsic viscosity of the sample(s) must be entered to display an<br />

adjusted calibration curve.<br />

Sample Data<br />

Use this area to input K and a or measured intrinsic viscosity values for<br />

your samples. These values in conjunction with the Mark-Houwink<br />

constants/intrinsic viscosity specified in the calibration table are used to<br />

adjust the calibration curve to better represent the sample being<br />

analyzed.