42 Setup and Control of Knauer HPLC Systems<br />

Fig. 53 LAN interface Device Information window<br />

Serial Number<br />

If you have read-out the device via LAN, the serial number has been<br />

added automatically. Otherwise type in the serial number of the pump<br />

you are using. The serial number of network controlled pumps is used for<br />

the device identification and addressing during communication. This<br />

serial number must match the serial number of the defined pump. If not,<br />

you will get an appropriate error message if you later try to start in run.<br />

Head<br />

Select the type of pump head you are using from the drop-down list. The<br />

pump head must also be selected in the pump’s own menu. For the<br />

pumps K-1001, K-1000, K-501 and K-120 pumps the choices are 10 ml<br />

and 50 ml of different materials (steel, PEEK, ceramics, titan).<br />

Fig. 54 Available pump heads for K-1001, K-1000, K-501, and K-120 pumps<br />

For the pumps S 1000, 1050, S 100 and 10P/20P/P2.1S/P4.1S only<br />

10 ml or 50 ml is to select, not considering the material. The pump P2.1L<br />

and the preparative pump 1800 can be equipped with four different pump<br />

heads with 100, 250, 500 or 1000 ml respectively. In case of titanium<br />

inlays also select the only available steel pump head for the preparative<br />

pump 1800.<br />

Fig. 55 Available pump heads S 1000 / 1050 / 100<br />

A B<br />

Fig. 56 Available pump heads P2.1L (A ); preparative pump 1800 (B)<br />

Once you have chosen the pump head, the related information about<br />

material, maximal flow and maximal pressure is displayed automatically.<br />

If you have read-out the device via LAN, the pump head type has been<br />

selected automatically.