114 Creating an Instrument Control Method<br />

Needle Wash<br />

Check the Needle Wash box to enable a wash procedure.<br />

Alarm Buzzer<br />

Check the Alarm Buzzer box to enable the alarm beep which is emitted<br />

when an error occurs during a run.<br />

Run Screens<br />

Check the Run Screens box to display the actual run information on the<br />

autosamplers display during a run.<br />

Temperature control (<br />

If your autosampler is equipped with a tray cooling and/or a facility<br />

thermo stated column oven (AS 3900 only) this options can be checked<br />

and the desired temperatures can be entered.<br />

The tray cooling allows to enter a temperature between + 4 °C and<br />

+ 15 °C.<br />

The AS 3900 oven can only control the temperature at ambient<br />

temperature + 5 °C and up. The maximum temperature is + 60 °C.<br />

Timed Events<br />

These types of events or methods enable you to control other devices via<br />

the two auxiliaries (contact closures) or to activate a new oven set point.<br />

Each event can be programmed on a time base, which starts at the<br />

moment of injection.<br />

Click the Timed events... button to set up the time base methods. The<br />

time base methods set up dialog will appear.<br />

Fig. 164 Autosampler timed events<br />

Click the New button to add one more lines to the programmed methods.<br />

Method type:<br />

Select one of the available AUX-1 and AUX-2. Every auxiliary AUX-1<br />

through AUX-2 can be set to ON (contact closure) four times and four<br />

times to OFF at a programmed Time. It is not possible to program the<br />

same auxiliary to ON and OFF at the same time.<br />

The maximal programmable time for the time base methods is 9 h 59<br />

min 59 sec.<br />

Click the Commit button to actualize your selection.<br />

Click the Delete button to remove the selected line from the programmed<br />

methods.