104 Creating an Instrument Control Method<br />

time program is similar in principal to that described above in the chapter<br />

Instrument Setup – Pumps, Pump Program.<br />

The Time, Excitation wavelength, and Emission wavelength values<br />

must be entered manually, whereas the options Autozero, Event Out,<br />

and Event Mark can be checked separately for each time line. Event Out<br />

short circuits the event out terminal for approximately 1 second. Event<br />

Mark adds an event mark to the analog output.<br />

The values for the Sensitivity, Gain, Recorder Range, and the Output<br />

Mode can be selected from the pull-down menus. Fields with selections<br />

available will display a combo-box button when the field or row is<br />

selected. Click the button to display the available choices.<br />

The available choices for the sensitivity are HIGH, MED, and LOW. F or<br />

the Output Mode the available choices are Excitation Energy, Light<br />

Source Correction, and Emission NOT corrected.<br />

Fig. 153 pull down menus in the Fluorescence Detector RF-10Axl setup<br />

Instrument Setup – Alltech 650 Conductivity Detector<br />

The Alltech 650 conductivity detector supports the general detector<br />

settings described on page 91.<br />

Fig. 154 Alltech 650 conductivity detector setup window<br />

Temperature Control<br />

Enter the desired temperature of the measuring cell, the allowed<br />

temperature tolerance and the waiting time for conditioning.<br />

Program table<br />

The program table can be edited in the usual way. For any time window<br />

you can select the range and the polarity of the signal via drop-down lists.<br />

The choice for the range is 0.01, 0.05, 0.10, 0.50, 1.00, 5.00, 10, 50, 100,<br />

500, 1000, and 5000 µS.<br />

Instrument Setup – User defined Detector<br />

For any user defined detector only the general detector settings,<br />

described on page 91, can be set.