30 Short guide ChromGate®, <strong>KNAUER</strong> Instrument Control<br />

Fig. 29 Instrument Status window, system status tab of a running method<br />

The running method is visualized by the dark dotted lines between the<br />

displayed instruments.<br />

For each pump, identified by name and serial number, the actual flow,<br />

pressure is displayed. If a gradient system is configured the solvent<br />

composition also will be shown.<br />

For each detector channel the wavelength and output values are shown.<br />

Similarly for all included instruments the relevant data will be displayed,<br />

e.g. the actual switching positions in case of the valves.<br />

All these information you also find on the single instrument tabs of this<br />

window, for details see pages 139 - 169. On these tabs you get the<br />

possibility of direct control interaction with the individual instruments. This<br />

is even possible during a method is running; however this option must be<br />

enabled while configuring the system (refer to Knauer instrument control<br />

method options on page 170).<br />

Shutting Down the System<br />

The system can be automatically shut down at the end of a sequence by<br />

including a shutdown method at the end of a sequence. A discussion of<br />

sequence types is presented in Chapter 4 of the reference manual.<br />

The shutdown method should include the following;<br />

� Column Oven temperature – value should be below ambient<br />

� Pump conditions, the flow for all pumps should be 0.0 ml/min and<br />

100% A and no events. For new pumps as Smartline pump 1050<br />

also a Standby option is available.<br />

� Deselect any timed events and Mix methods.<br />

� Set the Trigger to None. Now an autosampler will not be<br />

controlled and will not perform an injection.<br />

� Detector – activate the Lamp Off option or, if supported, the<br />

Standby option.